1. Functional check: Oscilloscope manual page 5.
Perform the functional check (photo).
First we performed an a Functional Check on the Oscilloscope as outlined on page 5 of the Oscilloscope Manual.
![]() |
| Testing a 10x Oscilloscope Probe as part of the Functional Check. |
2. Perform manual probe compensation (Oscilloscope
manual page 8) (Photo of overcompensation and proper compensation).
During the Functional check we discovered we found one of our probes was out of calibration, so we performed manual probe compensation in accordance with the Oscilloscope Manual Page 8.
Manual Probe Compensation Photos
![]() |
| Probe in a state of overcompensation. |
![]() |
| Probe reading after proper calibration. |
3. What does probe attenuation (1x vs 10x) do
(Oscilloscope manual page 9)?
Some probes have multiple attenuation settings that can be selected using a mechanical switch on handle of the probe. It is important to insure the setting you select matches the setting on the Oscilloscope. The attenuation settings change the impedance through the probes you are using, allowing a more accurate reading. If your probes settings do not match those of your Oscilloscope you will be unable to utilize the machine's full bandwidth potential.
4. How do vertical and horizontal controls work?
Why would you need it (Oscilloscope manual pages 34-35)?
Prior to starting this week's experiment we took some time to familiarize ourselves with the oscilloscope's controls. Our particular oscilloscope has two separate input channels and the control panel allows the user to choose whether to display data from either channel individually or both channels simultaneously.
The Horizontal and Vertical controls operate similarly to the controls on an old tube style television and allow you to control where the data is displayed on the screen. This is especially useful when displaying data from both channels as you can choose to either separate the data on different parts of the screen or directly overlay both channels data atop one another for easy comparison.
The Scale controls are also important as they allow the user to either expand or condense the data being displayed as needed to take proper reading.
Another important control is the Auto Set button which displays the scales the data on the display in an "optimized" manner with the press of a single button. While it is not always exactly what is called for it provides a good starting point for finer adjustment.
Also in conjunction with the Measure button the five buttons to the immediate right of the O-scope's screen allow the user to customize the information on the display.
5. Generate a 1 kHz, 1 Vpp around a DC 2 V from the
function generator (use the output connector). DO NOT USE oscilloscope probes for the function generator. There is
a separate BNC cable for the function generator.
a.
Connect this to the oscilloscope and verify the
input signal using the horizontal and vertical readings (photo).
b.
Figure out how to measure the signal properties
using menu buttons on the scope.
Our first task was to connect our function generator to our O-Scope and attempt to generate a 2VDC/RMS signal. Our instructions indicated we should have been able to accomplish this by generating 1 kHz / 1Vpp signal but even after increasing our output to amplitude to 2V (4Vpp) we were only able to achieve 1.50 V RMS
|
|
![]() |
| Function Generator putting out a 1 kHz 4Vpp Signal |
![]() |
| Despite increasing our input voltage up to 4Vpp we were only able to achieve 1.50 V RMS |
Note the RMS data displayed on the right hand side of the screen. The data displayed is customizable using the Menu/Measure buttons. In this case we selected Min/Max and Cycle/Cursor RMS Voltages.
6. Connect function generator and oscilloscope
probes switched (red to black, black to red). What happens? Why?
In accordance with the lab instructions we reversed the the black and red leads on the O-scope probe, while we expected to see a the waveform reverse, but instead the oscilloscope only displayed a flat zero volt signal.
We believe the reason for this that according to the O-scope manual (page v) the ground lead is at ground potential. Because of this the signal goes directly to ground and bypasses the oscilloscope's measurement circuitry.
7. After calibrating the second probe, implement the voltage divider circuit below. Measure the following voltages using the Oscilloscope and comment on your results:
&
8. For the same circuit above, measure Va and Vb using the handheld DMM both in AC and DC mode. What are your findings? Explain.
When attempting to implement the circuit below we were unable to achieve an input of .5 VAC/5 VDC utilizing our function generator.
We ended up using an input amplitude of 1.5 Volts with an offset of +3.72 Volts. When we attempted to either increase the offset or input amplitude the function generator gave a warning message.
The data we collected for Steps 7&8 is below
When taking our readings voltage division between the three resistors worked as expected with all three resistors dropping between 102-106 mV. The measured RMS voltage as measured by the O-Scope was almost exactly one -tenth of the DMM's VDC reading. While the DMM's AC reading was approximately equal to the O-Scope's V-RMS reading multiplied by sqrt(2).
9. For the circuit below
a. Calculate R so given voltage values are satisfied. Explain your work (video)
b. Construct the circuit and measure the values with the DMM and oscilloscope (video). Hint: 1kΩ cannot be probed directly by the scope. But R6 and R7 are in series and it does not matter which one is connected to the function generator.
| Step 9 Schematic Diagram |
We Calculated the Resistance using the following Steps
1.) Multiply 5V (RMS) by Sqrt 2 to Get an input Voltage of of 7.07 Vpp
2.) Calculate VR7 by subtracting VR6 from the 7.07V input. VR7=5.07 Vpp
3.) Calculate R7 using the equation R7 = R6(VR7/VR6)
4.) R7 = 2535 Ohms or 2.5 Kilo-Ohms
This is a short video describing how we calculated our R7 value for the given circuit.
This video chronicles the readings taken of circuit as requesting in step 9B, which confirmed the calculations we made in step 9A
10. Operational amplifier basics: Construct the
following circuits using the pin diagram of the Op-Amp. The half circle on top
of the pin diagram corresponds to the notch on the integrated circuit (IC).
Explanations of the pin numbers are below:
a.
Inverting amplifier: Rin = 1kΩ, Rf
= 5kΩ (do not forget -10 V and +10
V). Apply 1 Vpp @ 1kHz. Observe input and output at the same time.
What happens if you slowly increase the input voltage up to 5 V? Explain your
findings. (Video)
b.
Non-inverting amplifier: R1 = 1kΩ, R2
= 5kΩ (do not forget -10 V and +10
V). Apply 1 Vpp @ 1kHz. Observe input and output at the same time.
What happens if you slowly increase the input voltage up to 5 V? Explain your
findings. (Video)
Circuit 10 A (Inverting Amplifier) Schematic
Chris Explains the data we collected from the Inverting Amplifier.
A complete breakdown of this data can be found in the Table Below.
| Circuit 10 B Schematic |
A complete breakdown of this data can be found in the Table Below.